Operating & Maintenance ManualAutomatic Medical Gas ManifoldDome Loaded Regulator NFPA v1.2 www.amico.com
GAS SERVICE IDENTIFICATIONAmico manifolds are clearly labeled for the intended gas use, with the appropriate gas identied with a label that is attach
www.amico.com 11Display Functions1. Introduce power to the control cabinet. 2. Follow Cylinder Changing Procedures on the side of the cabin
12 Amico PipelineInstallationPRECAUTIONS• Tampering with gas specic connections shall be prohibited. Do not alter, remove or modify gas specic
www.amico.com 13Wall Mounting Instructions1. Remove manifold control panel from the shipping carton. If the manifold will be placed on a at
14 Amico Pipeline4. Hang the manifold control panel on the mounting bracket. The top two control panel mounting bolts will slide into slots of th
www.amico.com 15Cylinder Bank Installation Instructions Microprocessor Digital ManifoldPage: 4USER RESPONSIBILITYThe information contained in
16 Amico PipelineTesting for LeakageThe following instructions apply for performing a leak test on the joints made during assembly and connection
www.amico.com 17PrecautionsThe Amico Automatic Gas Manifolds are designed in accordance with the current revision of NFPA 99.There are three
18 Amico PipelineInitial Power-Up1. Release the two latches on either side of the manifold control panel and remove the cover. 2. Verify the fol
www.amico.com 1910. Slowly open one cylinder on the right side of the manifold.11. Verify the following:• Right bank Red “Empty” LED goes
User Responsibility 4Introduction 5Features 5Description of the Manifold 6Shipment Details 6The Manifold Enclosure 6Header Bar Wall Support Bracke
20 Amico PipelineHigh pressure gas is provided to the left and right manifold bank inlets via cylinders, pigtails and header assemblies. Flow of h
www.amico.com 21Manifold Switch OverAfter electrical power has been applied to the manifold, the side pressurized rst is designated the prim
22 Amico PipelinePerformance VericationUse the following test steps to verify the manifold’s functional performance:1. Remove manifold cover.2.
www.amico.com 23Performance Verication 14. Slightly depress the bleed valve on the left line regulator to create a small ow of gas through
24 Amico PipelineBank Regulator Pressure AdjustmentThis procedure should only be performed if the bank regulator pressures are not within acceptab
www.amico.com 25Should it be necessary to replace the bank regulator, it can be done while the manifold is in service. However, this should o
26 Amico PipelineLine Regulator Pressure AdjustmentThis procedure should be performed if the line regulator pressures are not within acceptable li
www.amico.com 27Control Board Replacement If necessary, the control board replacement can be performed while the manifold is in service. Howe
28 Amico PipelineCylinder Changing ProceduresWARNING: High pressure oxygen systems must be handled with CAUTION. Spontaneous combustion may occur
www.amico.com 29CONTROL CABINET PARTS LISTOrdering InformationThe Letters “XXX” Dene the Type of Gas:Oxygen = OXYNitrogen = NIT Medi
Cylinder Changing Procedures 28Ordering Information 29-30Control Cabinet Part List 29Appendix A: Manifold Internal Layout 31Appendix B: Piping Sch
30 Amico PipelineOrdering InformationDESCRIPTION MODEL NUMBERIntermediate check valve for all gasesOperating pressure relief valve NitrogenOperati
www.amico.com 31Appendix AManifold Internal Layout
32 Amico PipelineAppendix B6568743543671112VENT TOOUTSIDETOPIPELINEDISTRIBUTIONSYSTEMRIGHTBANKLEFTBANKHPLBHPRBRIGHT HAND BANKCYLINDER PRESSURELEFT
www.amico.com 33Appendix CWiring Schematic Heater UnitsThe Heaters normally switch on when the temperature drops below 24°C or 75°F. If the
34 Amico PipelineAppendix DElectrical Wiring DiagramAmico Medical Gas Alarm Master Alarm115 VACSupplyVoltageConnect the NO loop to the appropriate
www.amico.com 35Appendix EControl Cabinet Wiring DiagramLineSensorLine Display BoardLeft Bank Display Board Right Bank Display Board
36 Amico PipelineAppendix FStaggered Header Bar
www.amico.com 37Straight Header Bar SetupAppendix G
38 Amico PipelineAppendix HOptional Header Bar Setup
www.amico.com 39Standard Manifold Flow RatesAppendix I
4 Amico PipelineUser ResponsibilityThe information contained in this Installation and Maintenance Manual pertains only to the Amico Automatic Gas
Amico Pipeline | www.amico.com 85 Fulton Way, Richmond HillOntario, L4B 2N4, CanadaToll Free Tel: 1.877.462.6426Toll Free Fax: 1.866.440.4986Tel: 90
www.amico.com 5IntroductionThe Amico Automatic Gas Manifold is a one of a kind design that provides a reliable, uninterrupted supply of gas t
6 Amico PipelineDescription of the ManifoldSHIPMENT DETAILSThe Manifold system may be shipped in more than one carton, depending on the number of
www.amico.com 7Description of PartsWhen replacement cylinders are attached to the depleted bank, the Red LED goes out and the Yellow LED illu
8 Amico PipelineLINE REGULATOR BALL VALVEQuarter-turn ball valves are located upstream and downstream of each line regulator. These valves allow f
www.amico.com 9 Microprocessor Digital ManifoldPage: 4USER RESPONSIBILITYThe information contained in this Installation and Maintenance Manua